Structural comparison of dry and wet biochemistry analyzers | Seamaty

release time:2021-09-09 11:56:57

There are several chemical analysis techniques in biochemistry analyzers, including colorimetric reaction, ISE (ion selective electrode), immunoturbidimetry, etc. The way and container of colorimetric reaction is the main difference between dry and wet biochemistry analyzer.

The main difference between dry and wet biochemistry analyzers is the way and the vessel in which the colorimetric reaction occurs. Accordingly, the optical path systems of the two types of biochemical analyzers are also different.

Wet biochemistry analyzer

Normal wet biochemistry analyzers generally have color development reactions in a colorimetric cup or flow cell. Biochemistry analyzers that usually use a colorimetric cup are also equipped with a stirrer. This ensures an adequate and rapid reaction.

Therefore, a normal wet cell biochemistry analyzer must be equipped with a piping system for sample aspiration, sample dilution, reagent distribution and other functions.

Piping systems for flow cell biochemistry analyzers also need to perform functions such as pipeline cleaning and drainage of waste liquids. In addition, ordinary wet biochemical analyzers usually use photoelectric colorimetric method for analysis, and its absorbance and the concentration of the measured solution satisfy Lambert-Bier's law.

Dry biochemical analyzer

Dry biochemical analyzers, on the other hand, completely abandon the traditional pipeline system and use dry reagent strips as solid phase reagents. Dry reagent strips are generally composed of multiple layers of film.

Dry Biochemistry Analyzer

For example, some biochemical analyzers have a multilayer film composition of dry reagents as follows.
  1. The first layer is the distribution diffusion layer: it can make the specimen evenly distributed and can filter large molecules. Minimizes hemolysis, hyperlipidemia and bilirubin interference. Also provides a background for reflectometry.
  2. The second layer is the reagent layer: contains dry reagents. Controls the reaction sequence.
  3. The third layer is the indicator layer: contains the dyes to produce the color development complex.
The multilayer film technique not only masks the colored material to be measured and provides a background, selectively retains or removes interfering material. It also carries out a variety of substances in layers equivalent to the reaction principle of the wet chemical method.

The reaction products in one layer can in turn enter another layer for other reactions, thus directing the reaction sequence. Each layer can give a specific environment to control the reaction sequence and reaction time. The concentration of the components to be measured in the sample can be calculated by the Kubelka-Munk theory.

These are the structural differences between dry and wet biochemical analyzers.
Prev Return Next

Related Reading

If You Want to Know All About Clinical Chemistry Analyzer Read This One

2022-08-03

A clinical chemistry analyzer is a machine used in the diagnosis and treatment of disease. It is used to measure various chemicals in the blood, urine, or other body fluids. This information can be used to help identify and diagnose problems. Clinical chemistry analyzers are an important part of modern healthcare. They allow for fast and accurate analysis of patient samples. This helps ensure that patients receive the best possible care.

CBC Test Machine: the Effect of 7 Specimen Handling Methods on the Accuracy

2022-07-21

The cbc test machine plays an important role in clinical medical diagnosis. Inaccurate blood test results may mislead doctors to make wrong judgments and cause adverse consequences. The following are 7 ways in which the handling of blood samples may affect the accuracy of the cbc test machine.

Veterinary Blood Chemistry Analyzer Brands

2022-06-27

In order to determine the many components that makeup animal blood, a veterinary blood chemistry analyzer is used. Veterinary blood chemistry analyzers are utilized in healthcare facilities and in medical labs, forensic labs, and private homes.