Biochemistry Analyzer | Single wavelength and dual wavelength

release time:2021-10-09 15:13:33

In the instructions for biochemistry analyzers we find single wavelength and dual wavelength. What does this mean?

1. Single wavelength biochemical analyzer

The way of using one wavelength to detect the light absorption intensity of a substance is called single wavelength way. When the biochemical analyzer contains a component in the reaction solution. Or the absorption peak of the component to be measured in the mixed reaction solution, there is no overlap with the absorption wavelength of other co-existing substances. Single wavelength can be selected.

2. Dual wavelength biochemical analyzer

In the absorbance detection, the use of a primary wavelength and a secondary wavelength is called dual wavelength mode. If there is a large absorption of interfering substances in the reaction solution of the biochemical analyzer, affecting the accuracy of the measurement results. It is better to use the dual wavelength method.

Role of dual wavelength
  1. a) Eliminate noise interference;
  2. b) Reduce the effect of stray light;
  3. c) reduce the interference of light absorption of the sample itself.

The entire optical path system from the light source, to the colorimetric cup, monochromator, detector. There is an unstable detection signal that changes with time. That is, noise. The dual-wavelength detection is carried out simultaneously, the noise generated by the two wavelength detection is basically the same. Thus, noise interference can be eliminated.

When there are non-chemical interferents in the sample. Such as triglycerides, hemoglobin, bilirubin, etc. Will produce non-specific light absorption, and interfere with the accuracy of the measurement results. The use of dual wavelength mode of measurement can partially eliminate such interference and improve the accuracy of detection. 

3. The application of dual wavelength biochemical analyzer


For certain analytical items with fast reaction speed and cannot be set to two-point endpoint method. Especially in single reagent analysis, the dual wavelength method can be used to partially eliminate the light absorption interference of the sample itself.

The items that are measured by single reagent method and in which dual wavelength is applied are as follows.
  1. a) Total serum protein (bicuculline method) at a primary wavelength of 500 nm and a secondary wavelength of 576 nm.
  2. b) Serum albumin (bromocresol chloride method) primary and secondary wavelengths of 600 and 700 nm, respectively.
  3. c) Calcium (azoarsenic method) primary and secondary wavelengths of 660 and 770 nm, respectively.
  4. d) phosphorus (UV colorimetric method) primary and secondary wavelengths of 340, 405nm.
  5. e) magnesium (dimethylaniline blue method) primary and secondary wavelengths of 505 and 600nm.   
Prev Return Next

Related Reading

Key Indicators for Evaluating the Hematology and Urine Analyzers

2022-07-13

Since the birth of the world's first automatic blood cell counter in 1953, hematology analyzers have grown along with Laboratory Medicine. With the development of basic medicine and testing technology, and the application of computer technology, the detection ability of hematology analyzers has been continuously improved.

Clinical significance of biochemical indicators of liver damage in cats

2021-12-17

The liver, located at the diaphragm, is one of the most important organs in the abdominal cavity. It plays a vital role in maintaining the normal functioning and metabolism of the body, such as 1. aiding digestion (metabolism of fats) 2. synthesis of proteins and hormones.

Effect of pet diet, mood, and exercise on biochemical results in pets | Animal diagnosis

2021-10-26

Diet affects a variety of substances in the blood. For example, the concentrations of glucose (GLU), triacylglycerol (TG), ALP and phosphorus. Eating a high-protein diet one day before the blood draw can result in high urea nitrogen (BUN) and uric acid (UA) results. The increase in lipid concentrations in the blood, especially TG, after eating can lead to a milky cloudy serum, which can interfere with biochemical measurements. This may result in high results for bile acids, proteins, calcium and phosphorus. The results of amylase measurements are low.