Why Should a Biochemical Analyzer Be Calibrated?

release time:2022-04-11 15:24:44


1. The meaning of biochemistry analyser calibration
The biochemical analyzer calibration is to find a reference point, which is a K value (or F value). It is determined by the state of the biochemical instrument and reagents. When we measure a specimen, whether you use a manual method or a fully automatic biochemistry analyzer, the value measured is just an absorbance.

This absorbance means nothing to us. We have to convert this absorbance into a concentration or enzyme activity. Then we have to multiply it by a K value. The K value is what we find by calibrating. The general minimum requirement is to have a reagent blank and a standard, after the instrument has determined two absorbances.

K = (standard concentration - reagent blank) / (A standard - A reagent blank)
(the reagent blank is usually 0)

The concentration of the standard is known to us. These two absorbances can be measured by the biochemistry machine. This gives a K value. Whatever the specimen, multiplying its absorbance by the K value gives us the answer. The K value is therefore very decisive and can determine the accuracy of a specimen.
Biochemical Analyzer Calibrate
2. biochemistry machine K-value determinants

Let's look at what affects the K-value. Firstly, the concentration of the standard solution must be accurate (it is best to use the same serum as the specimen). Secondly, the absorbance of the standard must be accurate in relation to the absorbance of the reagent blank. The absorbance is influenced by the condition of the biochemistry machine and the condition of the reagents. If your biochemistry machine is stable, the reagents are a major factor in the K-value.

3. How can I be sure that the biochemical analyzer K-value is correct?

We generally check this with a QC serum. It is best to have two levels of QC sera to check. If the QC results are good, it can be said that the K value is accurate. The patient's result calculated using this K value is also accurate, so the K value is very critical.

4. The true nature of the biochemical analyzer K-value

The K value actually represents the slope and the intercept represents the reagent blank, which changes every day. If the biochemistry machine and reagents are stable, then the K value is also stable.

5. How often is the calibration appropriate?

This depends on the stability of the reagents. If the quality control results are not good, some projects can be solved by doing a reagent blank, while some projects need to do two-point calibration.

6. Does a daily calibration guarantee stable biochemistry machine results?

Not necessarily, the calibration may not be freshly dissolved every day. Some items are not stable after re-dissolution of the calibration solution, such as TBIL, DBIL (decomposition by light), GLU (bacterial decomposition) and enzyme items (degradation by re-dissolution after freezing). Solution: The above items must be freshly dissolved in a bottle of calibrator when calibrating, at least for 30 minutes, and measured within 1 hour.

The above is the problem related to the calibration of the biochemical analyzer. I hope you find it helpful.
Prev Return Next

Related Reading

Diagnostic Devices for IVD-Medical Equipment and Supplies

2022-10-10

IVD instruments can be divided into clinical chemical analysis instruments, immunochemical analysis instruments, blood analysis instruments and microbiological analysis instruments if according to the diagnostic methods and objects.

13 factors affecting biochemical analyzer test results

2021-10-14

Biochemical analyzers can be divided into fully automatic biochemical analyzers and semi-automatic biochemical analyzers. Nowadays, most hospitals use fully automatic biochemical analyzers.

Structural comparison of dry and wet biochemistry analyzers | Seamaty

2021-09-09

There are several chemical analysis techniques in biochemistry analyzers, including colorimetric reaction, ISE (ion selective electrode), immunoturbidimetry, etc. The way and container of colorimetric reaction is the main difference between dry and wet biochemistry analyzer.